Dot-x SERIES Especificações Página 47

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 46
DOTX-45 - 46 -
Part 7 Calibration
7.1 Oxygen Calibration
Once power is applied, the sensor must be given time to stabilize. This is best
done by following the zeroing procedure below. Establishing a stable zero is
critical to the proper operation of the monitor. A complete calibration will include
zeroing and spanning the sensor. It is generally unnecessary to set the zero at
every calibration, however, it should be done during the initial installation.
7.11 Oxygen Span Cal
The system provides three methods of D.O. calibration: 1-Point (sample), and
% Saturation (air cal), and Zero. These three methods are significantly different.
It is important to note that ONLY the Air Calibration method outlined in
section 7.13 may be used to calibrate a D.O. monitor that has been
programmed as Instrument Type 3 in the configuration menu. This type
of instrument is designed to measure % Saturation and must be calibrated
in air using ambient air as a 100% saturation standard.
7.12 Dissolved Oxygen Span Cal (1-spl)
The 1-Point (sample or comparison) method is intended to be primarily used as
an on-line calibration; however, the sensor can be removed, cleaned and then
calibrated in a bucket of clean water if necessary. During calibration, the system
will display the current ppm reading and the user can manually enter a reference
value from a lab sample or comparative reference instrument. In the DOTX-45
system, the 1-Point calibration adjusts the slope of the sensor output response.
1. Determine whether the calibration will be done on-line or with the sensor
removed and placed into a bucket of clean water. If the sensor is removed
from the application, rinse and clean if necessary.
2. If the sensor has been removed and placed into a bucket of water, allow
sensor to temperature equilibrate with the solution as much as possible. With
the sensor coming from an application which differs greatly in temperature,
the user may have to wait as much as 20 minutes. If the sensor is on-line, the
user may want to set the output HOLD feature prior to calibration to lock out
any output fluctuations.
3. Scroll to the CAL menu section using the MENU key and press ENTER or the
UP arrow key. Cal DO will then be displayed.
Vista de página 46
1 2 ... 42 43 44 45 46 47 48 49 50 51 52 ... 68 69

Comentários a estes Manuais

Sem comentários